EC Sensor, Hydrogen, PEEK, 50 bar, O-Ring FKM/FPM

EC Sensor, Hydrogen, PEEK, 50 bar, O-Ring FKM/FPM
Artikel-Nr.: 31210.12
Einzelpreis Kontakt

Inhalt

Sensor and manual.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.

Technische Daten

Anströmgeschwindigkeit: Typical Flow Rate
Membrane 2956A : 50 to 220 mL/min
Membrane 2952A : 40 to 200 mL/min
Membrane 2995A : 20 to 70 mL/min
Membrane 29015A : 20 to 40 mL/min
Anwendung: Corrosion rate
Genauigkeit: The greater of ±1% of reading, or ±0.03 ppb, or ±1 Pa (Membrane 2956A)
  The greater of ±1% of reading, or ±0.09 ppb, or ±6 Pa (Membrane 2952A)
  The greater of ±1% of reading, or ±10 ppb, or ±1 kPa (Membrane 29015A)
  The greater of ±1% of reading, or ±1 ppb, or ±50 Pa (Membrane 2995A)
Gewährleistung: 24 Monate
Gewicht: 140 to 700 grams, depending on the construction materia
Inhalt: Sensor and manual.
Controller, sensor cable, flow-chamber or process connection must be ordered separately.
Material: O-Ring: FKM/FPM
Messbereich: Range at 25 °C Membrane 2956A : 0 - 75 ppb or 0 Pa - 5 kPa
Membrane 2952A : 0 - 300 ppb or 0 Pa - 20 kPa
Membrane 2995A : 0 - 3200 ppb or 0 Pa - 200 kPa
Membrane 29015A : 0 ppb - 32 ppm or 0 - 2000 kPa
Parameter: Hydrogen
Probendruck: 0 - 50 bar
Schutzklasse: IP68 / NEMA4
Spezielles Ausstattungsmerkmal: Silver guard ring
Strahlendosis-Grenzwert: Membrane 2956A : 2x10⁴ rad
Membrane 2952A : 108 rad
Membrane 2995A : 108 rad
Membrane 29015A : N/A
Temperaturbereich: Accurate from -5 - 50 °C (23 - 122 °F)

Resistant - temperature from -5 - 100 °C (23 - 212 °F)
Umgebungstemperatur: -5 - 100 °C (23 - 212 °F)
Zertifizierung: 2004/108/EC - EN 61326-1