GLI Model C33 Contacting Conductivity Controller

VERALTETER ARTIKEL

GLI Model C33 Contacting Conductivity Controller
Artikel-Nr.: C33A1NN

VERALTETER ARTIKEL

Multiple Measurements

Sensor A and B can be independently selected to measure conductivity (μS/cm or mS/cm), resistivity (MΩ • cm or KΩ • cm), or total dissolved solids (ppm or ppb). The measured values or temperatures for Sensor A or B can be displayed separately or together (four total measurements). Both analog outputs can also be displayed. The Model C33 can provide a calculated Sensor A and B measurement (% rejection, % passage, ratio A/B, ratio B/A, difference A-B, or difference B-A). Front panel red LEDs indicate relay on/off status.

Ultra-pure Water Capability

Model C33 accurately measures conductivity down to 0.056 μS/cm and resistivity up to 18.2 MΩ • cm, both at 25.0°C.

Backlit LCD Readout

The two-line, 16 character backlit LCD provides excellent viewing contrast under all lighting conditions.

Menu-guided Operation

The simple keypad and logical menu structure make this analyzer easy to use. Menu screens guide you through setup, calibration, operation, and test/maintenance functions.

Multiple Language Capability

All screens can be selected for display in English, French, German, or Spanish.

Inhalt

Controller, User Manual

Technische Daten

Alarm: Alarm: settings for low alarm point, low alarm point deadband, high alarm point, high alarm point deadband, off delay, and on delay
Anzeige: 2-line by 16 character backlit LCD
Bemerkungen: Calculated sensor A and B measurement:
0-100% rejection
0-100% passage
ratio A/B or B/A
difference A-B or B-A
Besondere Hinweise: Relay A and B LEDs indicate respective relay status
Betriebsbedingungen: -20 - 60 °C at 0 - 95% RH (non-condensing)
Datenspeicher: All user settings are retained indefinitely in memory (EEPROM) non-volatile
Drift: 0.05 % of span per 24 hrs., non-cumulative
Energieverbrauch: 10 VA
Genauigkeit: 0.1 % of span
Gewährleistung: 24 Monate
Gewicht: 0,8 kg
Höhe: 104 mm
Inhalt: Controller, User Manual
Kalibriermethode: GLI DRY-CAL Method:
enter the sensor’s GLI-certified “calibration constant” and temperature “T” factor
1-POINT SAMPLE Method:
enter one sample value derived by laboratory analysis or comparison reading
ZERO Method:
With the dry sensor in air, press keys to initiate automatic system zeroing
Kanäle: 2 sensor A or B’s measured values or temperatures can be separately displayed, or all four measurements can be displayed together on
one screen. Both outputs are displayed together on a screen. Also, a calculated sensor A and B measurement can be displayed
Lagerbedingungen: -30 °C - 70 °C
Material: Brackets: Zinc-plated steel
Max. Kabellänge von Sonde zu Controller: 91 m
Messbereich: -20 - 200 °C
  selectable:
0-9999 ppm
0-9999 ppb
Messbereich LF: Selectable:
0-2.000
0-20.00
0-200.0
0-2000
(µS/cm or mS/cm)
Montage: Panel mounting
Relais: 2 electromechanical relays; SPDT (Form C) contacts; 5A 115/230 V AC, 5A at 30 V DC resistive
Relais Funktionen: Control: settings for high/low phasing, setpoint, deadband, overfeed timer, off delay, and on delay
Status: not configurable; relay only activates when a "fail" diagnostic condition exists
Relative Luftfeuchtigkeit: 0 - 95 % non-condensing
Relay: Signalquelle: Each relay (A and B) can be assigned to be driven by
sensor A's or sensor's B selected measurement (conductivity, resistivity or TDS) or temperature
Schnittstelle: RS-232
Spannungsversorgung (Hz): 50 - 60 Hz
Spannungsversorgung (Volt): 180 - 260 V AC
Stromausgang: 2 x 0/4 to 20 mA each with 0.004 mA (12-bit) resolution and capability to drive up to 600 Ω loads
Temperatur-Abweichung: < 0.03 % of span per °C (zero and span)
Temperatur-Kompensation: -20 - 200 °C automatic or manual with selection for temperature compensator (Pt1000 RTD or Pt100 RTD); Linear % per °C slope, built-in ammonia temperature properties table, built-in natural water temperature properties table, or no compensation
  to automatic or manual with selection for temperature compensator (Pt1000 RTD or Pt100 RTD); Linear % per °C slope, built-in ammonia temperature properties table, built-in natural water temperature properties table, or no compensation
Wiederholbarkeit: 0.1 % of span or better